MAUDE Adverse Event Report: AIZU OLYMPUS CO., LTD. EVIS LUCERA ELITE COLONOVIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AIZU OLYMPUS CO., LTD. EVIS LUCERA ELITE COLONOVIDEOSCOPE Back to Search Results
Model Number PCF-H290DL
Device Problems No Display/Image (1183); Image Display Error/Artifact (1304)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/21/2024
Event Type  malfunction  
Manufacturer Narrative
A review of device history record and historical complaints analysis was conducted and no deviations that could have caused or contributed to the reported issue was identified.The device was returned, and the evaluation found no additional reportable malfunctions other than those mentioned in b5.During initial inspection of the device, error code e315 unsupported scope was evident.Image was not present when the endoscope was connected and switched on.Error code e315 was displayed on the screen, the plug body was dismantled, the moisture indicator had been triggered.Quality trend analysis has shown that fluid inside the plug body can result in intermittent image issues temporarily affecting the video image and other electrical functionality.The most likely cause of fluid ingress is during the manual leak check or cleaning process and is therefore typically attributed to user handling.The suggested event is detectable and preventable by handling the scope in accordance with the following sections of the instructions for use: important information ¿ please read before use ¦precautions: caution turn the video system center on only when the endoscope connector is connected to the light source.In particular, confirm that the video system center is off before connecting or disconnecting the endoscope connector.Failure to do so can result in equipment damage, including destruction of the image sensor.¦precautions for disappeared or frozen endoscopic image: warning follow the precautions given below.Otherwise, the endoscopic image may disappear unexpectedly or the frozen image may not be restored during the examination.3.8 inspection of the endoscopic system ¦inspection of the endoscopic image confirm that the wli and nbi endoscopic images (except bf-mp290f) are normal.5.1 troubleshooting if any irregularity is observed during the inspection described in chapter 3, ¿preparation and inspection¿, do not use the endoscope and solve the problem as described in section 5.2,¿troubleshooting guide¿.If the problem still cannot be resolved, send the endoscope to olympus for repair as described in section 5.4, ¿returning the endoscope for repair¿.Also, should any irregularity be observed while using the endoscope, stop using it immediately and withdraw the endoscope from the patient as described section 5.3, ¿withdrawal of the endoscope with an irregularity¿.Olympus will continue to monitor the field performance of this device.
 
Event Description
It was observed that during the device evaluation, the colonovideoscope exhibited error code 315 unsupported scope and compromised image.There were no reports of patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS LUCERA ELITE COLONOVIDEOSCOPE
Type of Device
COLONOVIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA   965-8520
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18923452
MDR Text Key337896838
Report Number9610595-2024-05705
Device Sequence Number1
Product Code FDF
Combination Product (y/n)N
Reporter Country CodeUK
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 03/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/18/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberPCF-H290DL
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/26/2024
Was the Report Sent to FDA? No
Date Manufacturer Received02/26/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/05/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-