BAXTER HEALTHCARE CORPORATION FOLFUSOR; PUMP, INFUSION, ELASTOMERIC Back to Search Results
Catalog Number 2C4702K
Device Problem No Flow (2991)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was reported that a small volume folfusor had no flow; further described as "the pump contained the same volume of solution after two days".This was observed during patient use.The device contained 96ml 5-fluorouracil.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
E1: initial reporter facility name: (b)(6)/ e1: initial reporter city: (b)(6).E1: initial reporter postal code: (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
Additional information: d4, d9, h3, h4, h6, h10 h4: the lot was manufactured between march 8, 2023 - march 9, 2023 h10: the actual device was received for evaluation containing 94ml of fluid in the bladder.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.After the luer cap was removed, evidence of continuous flow of fluid was observed flowing out of the distal luer.A functional flow rate test was performed, and the flow rates were found to be within the product specification range.During the flow test, evidence of continuous flow of fluid was observed flowing out of the distal luer.Based on the evaluation result, the folfusor unit was determined to be conforming product.The reported condition was not verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FOLFUSOR
Type of Device
PUMP, INFUSION, ELASTOMERIC
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - IRVINE
17511 armstrong avenue
building 3
irvine CA 92614
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key18923466
MDR Text Key337896601
Report Number1416980-2024-01171
Device Sequence Number1
Product Code MEB
UDI-Device Identifier00085412080154
UDI-Public(01)00085412080154
Combination Product (y/n)N
Reporter Country CodeSZ
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Pharmacist
Type of Report Initial
Report Date 03/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/18/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number2C4702K
Device Lot Number23C012
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/21/2024
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
5-FLUOROURACIL
-
-