It was reported that the device was in backup mode and was unable to be interrogated via radiofrequency telemetry.The device was explanted and replaced to resolve the event and the patient was in stable condition.The device is included in the zenex, assurity, endurity laser adhesion preparation advisory issued by abbott on 20 july 2022 for a subset of devices distributed and implanted outside of the united states.
|
The reported event of no rf telemetry was confirmed.The reported event of device in backup was not confirmed.The device had no output and no telemetry.Visual inspection of the header attachment area detected an anomaly between the pre-molded header and titanium case.The device was cut open to enable further testing and the battery was found at normal levels.Feedthrough leak test was performed, indicating a device hermeticity breach.This is consistent with feedthrough damage as a result of fluid intrusion between the header and case, and subsequent fluid ingress to internal electronics.Hybrid circuitry was tested by connecting to an external power source.Test results indicated normal current drain.A manufacturing process anomaly consistent with header bonding anomaly may have occurred.The device is included in the zenex, assurity, endurity laser adhesion preparation advisory issued by abbott on 20 july 2022 for a subset of devices distributed and implanted outside of the united states.
|