BOSTON SCIENTIFIC CORPORATION TRAPEZOID RX; LITHOTRIPTOR, BILIARY MECHANICAL Back to Search Results
Model Number M00510870
Device Problem Detachment of Device or Device Component (2907)
Patient Problem Foreign Body In Patient (2687)
Event Date 02/19/2024
Event Type  malfunction  
Event Description
It was reported to boston scientific corporation that a trapezoid rx was used in the ductus hepaticus (hepatic duct) during a lithotripsy procedure performed on (b)(6) 2024.During the procedure, the basket was used in an attempt to remove a bile stone.While trying to crush the stone using the lithotripsy handle, the whole basket fell off, and it was found in the duodenum using fluoroscopy.It is currently unknown if the basket was removed during the procedure or if it was planned to be removed.A different device was used to complete the procedure.There were no patient complications as a result of the event.
 
Manufacturer Narrative
Block h6: imdrf device code a0501 captures the reportable event of basket detachment.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRAPEZOID RX
Type of Device
LITHOTRIPTOR, BILIARY MECHANICAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
780 brookside drive
spencer IN 47460
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key18924940
MDR Text Key337917585
Report Number3005099803-2024-00989
Device Sequence Number1
Product Code LQC
UDI-Device Identifier08714729296386
UDI-Public08714729296386
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K040447
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative
Reporter Occupation Nurse
Type of Report Initial
Report Date 03/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/18/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM00510870
Device Catalogue Number1087
Device Lot Number0032106229
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/14/2024
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/21/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured07/28/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-