• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOSENSE WEBSTER INC THERMOCOOL® SMART TOUCH® SF BI-DIRECTIONAL NAVIGATION CATHETER; SIMILAR DEVICE D132701, PMA # P030031/S053

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOSENSE WEBSTER INC THERMOCOOL® SMART TOUCH® SF BI-DIRECTIONAL NAVIGATION CATHETER; SIMILAR DEVICE D132701, PMA # P030031/S053 Back to Search Results
Catalog Number D134825IL
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/02/2024
Event Type  malfunction  
Event Description
It was reported that a patient underwent an atrial fibrillation (afib) - paroxysmal ablation procedure with a thermocool® smart touch® sf bi-directional navigation catheter for which biosense webster¿s product analysis lab (pal) identified a reddish material and a hole in the pebax.Initially, it was reported that during the operation, there was a problem with the force of the catheter.A second device was used to complete the operation.There was no adverse event reported on patient.The force issue was assessed as non mdr reportable.The potential risk that it could cause or contribute to a serious injury or death to the operator or patient was remote.The biosense webster, inc.Product analysis lab received the device for evaluation and per the evaluation completion on 21-feb-2024, there was a reddish material and a hole in the pebax.This was assessed as mdr reportable with an awareness date of 21-feb-2024.
 
Manufacturer Narrative
The thermocool® smart touch® sf bi-directional navigation catheter was returned to biosense webster (bwi) for evaluation.A visual inspection and screening test of the returned device were performed following bwi procedures.Visual analysis of the returned sample revealed reddish material and a hole in the pebax.This could be related to the handling and/or manipulation during the procedure.However, this cannot be conclusively determined.The magnetic and force features were tested, and no errors were observed.The force values and the vector were observed within specifications.No force issues were observed.The blood found inside the pebax area may have contributed to the force and magnetic issue.A manufacturing record evaluation was performed for the finished 30991634l device number, and no internal actions related to the reported complaint condition were identified.The issue reported by the customer was confirmed.The instructions for use contain the following information that should be considered: to ensure accurate force readings, verify that the force reading is near zero when the catheter is not in contact with tissue.If the force reading is not near zero when the catheter is not in contact with tissue, perform zeroing.As part of biosense webster¿s quality process, all devices are manufactured, inspected, and released to approved specifications.The ¿suspected medical device¿ reported in section d of this report is not marketed in usa or approved by the fda.However, it is being reported as biosense webster considers this as a similar device to thermocool® smart touch® sf bi-directional navigation catheter approved under pma # p030031/s053 this report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster, inc., or its employees that the report constitutes an admission that the product, biosense webster, inc., or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Manufacturer's reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THERMOCOOL® SMART TOUCH® SF BI-DIRECTIONAL NAVIGATION CATHETER
Type of Device
SIMILAR DEVICE D132701, PMA # P030031/S053
Manufacturer (Section D)
BIOSENSE WEBSTER INC
31 technology drive, suite 200
irvine CA 92618
Manufacturer (Section G)
BIOSENSE WEBSTER INC (JUAREZ)
circuito interior norte
1820parque industrial salvacar
juarez 32599
MX   32599
Manufacturer Contact
kate karberg
31 technology dr
irvine, CA 92618
3035526892
MDR Report Key18925703
MDR Text Key337956266
Report Number2029046-2024-00891
Device Sequence Number1
Product Code LPB
Combination Product (y/n)N
Reporter Country CodeCH
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other
Type of Report Initial
Report Date 03/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/18/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/31/2024
Device Catalogue NumberD134825IL
Device Lot Number30991634L
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/14/2024
Date Manufacturer Received02/21/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/01/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-