SMITHS MEDICAL ASD, INC. LEVEL 1 HOTLINE BLOOD AND FLUID WARMER; WARMER, THERMAL, INFUSION FLUID
|
Back to Search Results |
|
Model Number HL-90 |
Device Problem
Defective Alarm (1014)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/01/2024 |
Event Type
malfunction
|
Event Description
|
It was reported that the device is alarming that fluid warming set is not properly installed despite it being firmly installed.It was unknown when the fault occurred.It was brought to the biomed department at the start of the day with a note saying, "not working".No patient or clinical injury was reported.
|
|
Manufacturer Narrative
|
B3: date of event is unknown; no information has been provided to date.Device evaluation: one device was received.Per visual inspection, enclosure cracked at quick connect, faded line cord, reservoir is worn out, and corroded quick connect.During functional testing the complaint was confirmed.The root cause was a faulty microswitch.It was unknown what caused it to become faulty.The product's history records were reviewed and there were no non-conformances nor service-related issues that would have resulted in the reported complaint.No action taken due to the condition of the device.It was deemed beyond economical repair and was scrapped.
|
|