Model Number HL-90
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/01/2024
Event Type  malfunction  
Event Description
It was reported that the reservoir is leaking.There was no patient involvement and no patient harm/adverse event reported.
 
Manufacturer Narrative
B3: date of event is unknown; no information has been provided to date.Device evaluation: one device was received.Per visual inspection, stained water tank, reservoir gasket is worn out, outdated printed circuit board (pcb), power switch and quick connect.Per functional testing, the device was leaking from the bottom of the water tank cover.The complaint was confirmed.The root cause was a worn gasket.It was unknown what caused it to become worn.The product's history records were reviewed and there were no non-conformances nor service-related issues that would have resulted in the reported complaint.No action taken due to the condition of the device.It was deemed beyond economical repair and was scrapped.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE BLOOD AND FLUID WARMER
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan ln n
minneapolis MN 55442
Manufacturer Contact
reed covert
6000 nathan lane north
minneapolis, MN 55442
2247062300
MDR Report Key18926665
MDR Text Key337957906
Report Number3012307300-2024-01448
Device Sequence Number1
Product Code LGZ
UDI-Device Identifier50695085820909
UDI-Public(01)50695085820909(11)000208
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K001764
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 03/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/18/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberHL-90
Device Catalogue NumberCON-HL-90
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/29/2024
Date Manufacturer Received02/20/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/08/2000
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-