Model Number 7122Q/65
Device Problems Difficult to Fold, Unfold or Collapse (1254); High impedance (1291); Capturing Problem (2891)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/27/2024
Event Type  malfunction  
Event Description
Related manufacturer reference number: 2017865-2024-35275; it was reported that the patient presented for implant procedure.During procedure it was noted that the right ventricular (rv) lead was exhibiting high capture threshold, high pacing impedance, and the helix had difficulty extending.The procedure was completed by replacing the rv lead.Following the procedure, the rv lead exhibited poor impedance and capture threshold.While attempting to reposition the rv lead, the helix had difficulty extending.The rv lead was successfully repositioned on 29 feb 2024.The patient was in stable condition.
 
Manufacturer Narrative
A device history record (dhr) review was performed and all required manufacturing processes and inspections steps were confirmed to be completed per the requirements.The device met specifications prior to leaving abbott manufacturing facilities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR
Type of Device
Permanent defibrillator electrodes
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18927333
MDR Text Key337956515
Report Number2017865-2024-35274
Device Sequence Number1
Product Code NVY
UDI-Device Identifier05414734503341
UDI-Public05414734503341
Combination Product (y/n)Y
Reporter Country CodeBR
PMA/PMN Number
P950022
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 03/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/18/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number7122Q/65
Device Lot NumberS000091017
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/27/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/28/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-