|
Model Number PACE 203H |
Device Problem
Failure to Sense (1559)
|
Patient Problem
Cardiac Arrest (1762)
|
Event Date 01/17/2024 |
Event Type
malfunction
|
Event Description
|
It was reported by the health professional: ventricular sensitivity measured at 7mv.Device set to 3mv.Despite this, after a period of pacing, the device failed to detect three r waves and delivered a pacing spike during repolarisation (t wave), triggering vf.
|
|
Manufacturer Narrative
|
The pacemaker with s/n (b)(6) has been tested first by the uk distribution partner dot medical who found it working as expected with now sign of sensing problems.The device has been forwarded to osypka medical, the manufacturer who received it on 23-feb-2024.The initial functional test did not discover any malfunction except some mechanical damages which are probably not related to the reported incident.Following two long term testing were performed: one 48h monitoring under temperature cycling condition and another 24h continuously sensing monitoring evaluation (using the parameter from the incident).The device has passed all tests without any findings: all parameters worked within its specification.No sensing problem could be observed.The initial reporter has been contacted asking for more ecg readings recorded 30 sec.Before the incident happened.This could not be provided, unfortunately.The root cause of the reported sensing problems could not be identified.If r-on-t pacing happened it could have been caused by the following reasons other than device malfunction: undersensing: sensing problems due to lead connection problems caused pacemaker switched to asynchronous pacing: pacemaker switched to asynchronous pacing due to the presence of interferences.This is a common safety feature on external pacemakers providing misinterpretation of noise signals as intrinsic activities.
|
|
Search Alerts/Recalls
|
|
|