MEDTRONIC IRELAND MICRA AV; LEADLESS PACEMAKER Back to Search Results
Model Number MC1AVR1
Device Problems Failure to Capture (1081); High impedance (1291); Pacing Problem (1439); Material Integrity Problem (2978); Unstable Capture Threshold (3269)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/07/2024
Event Type  malfunction  
Event Description
It was reported intra operatively that the leadless implantable pulse generator (ipg) exhibited high impedance values due to tissue being attached to it, pacing malfunction and unstable thresholds.It was also noted that the delivery system exhibited a hard and would not move smoothly and that the tether may be broken.The leadless ipg and delivery system were attempted/not used and replaced. no patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Manufacturer Narrative
Corrected b5 (second sentence) and h6: fdd code medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event Description
It was reported intra operatively that the leadless implantable pulse generator (ipg) exhibited high impedance values due to tissue being attached to it, pacing malfunction and unstable thresholds.It was also noted that the delivery system was hard and would not move smoothly.The leadless ipg and delivery system were attempted/not used and replaced. no patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product event summary: the device was returned and analyzed.Returned product analysis was performed and no anomalies were found.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MICRA AV
Type of Device
LEADLESS PACEMAKER
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
EI 
Manufacturer (Section G)
MEDTRONIC IRELAND
parkmore business park west
galway
EI  
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key18932047
MDR Text Key338452335
Report Number9612164-2024-01312
Device Sequence Number1
Product Code PNJ
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
P150033
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 03/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/19/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMC1AVR1
Device Catalogue NumberMC1AVR1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/22/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured07/19/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age80 YR
Patient SexMale
-
-