BOSTON SCIENTIFIC CORPORATION TACTRA MALLEABLE PENILE PROTHESIS; PROSTHESIS PENILE Back to Search Results
Model Number 720080-01
Device Problem Migration (4003)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/27/2024
Event Type  Injury  
Event Description
It was reported that the patient with this tactra penile prosthesis presented with cross over of the implant.A ct scan determined that the device had crossed over towards the proximal end.The physician did not suspect the cross over was due to the device, but rather the sub coronal approach not providing easy proximal dilation.The device was explanted, and a new tactra was implanted.There were no patient complications reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TACTRA MALLEABLE PENILE PROTHESIS
Type of Device
PROSTHESIS PENILE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
Manufacturer Contact
farshad fahimi
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key18933178
MDR Text Key338025870
Report Number2124215-2024-16266
Device Sequence Number1
Product Code FAE
UDI-Device Identifier08714729979340
UDI-Public08714729979340
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K183619
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Study,Health Professional
Reporter Occupation Physician
Type of Report Initial
Report Date 03/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/19/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number720080-01
Device Catalogue Number720080-01
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/28/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Hospitalization;
Patient Age72 YR
Patient SexMale
-
-