Reportable based on device analysis completed on (b)(6) 2024.It was reported that when the first coil was deployed, it felt like it was in a bad position and could not be withdrawn.The 80% stenosed target lesion was located in the mildly tortuous and moderately calcified pulmonary artery.An 8mm x 20cm interlock-35 was selected for use.During the procedure, when the first coil was deployed, it felt like it was in a bad position, and it could not be withdrawn.The coil was withdrawn with the catheter.The procedure was completed with another device.No complications were reported, and patient was stable post procedure.However, device analysis revealed that the zap tip was detached.
|
E1; initial reporter facility name - beijing anzhen hospital, capital medical university device evaluated by mfr: the device was returned for evaluation.It was observed that the main coil was bent and stretched at the coil arm and primary coil section.Also, the zap tip was detached.The functional could not be performed due the main coil and the pusher wire are not interlocking.Dimensional inspection of the pusher wire and main coil revealed the components were within specifications.No other issues were identified during the product analysis.
|