DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, FRENCH, 220V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-3023-64
Device Problem Component Missing (2306)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/11/2024
Event Type  malfunction  
Manufacturer Narrative
A supplemental report will be submitted upon completion of our investigation.
 
Event Description
It was reported that the cs300 intra-aortic balloon pump (iabp) had no tightening system.The ready console arrived at the customer without clamping system or helium bottle.There was no patient involved.
 
Manufacturer Narrative
Updated data: b4, e1(initial reporter), e2, e3, g2, g3, g6, h2, h10.
 
Event Description
N/a.
 
Manufacturer Narrative
The field service engineer evaluated the unit and replaced the clamping system knob and tank valve.All functional and safety checks have been performed to meet factory specifications.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS300 INTRA-AORTIC BALLOON PUMP, FRENCH, 220V
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key18933381
MDR Text Key338101366
Report Number2249723-2024-01163
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108308
UDI-Public10607567108308
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K063525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date 05/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/19/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0998-00-3023-64
Device Catalogue Number0998-00-3023-64
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/02/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/12/2012
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-