• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC. LEVEL 1 HOTLINE FLUID WARMER, 390; WARMER, THERMAL, INFUSION FLUID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC. LEVEL 1 HOTLINE FLUID WARMER, 390; WARMER, THERMAL, INFUSION FLUID Back to Search Results
Model Number HL-390
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/01/2024
Event Type  malfunction  
Manufacturer Narrative
B3: date of event is unknown; no information has been provided to date.One device was received.Per visual inspection, the water tank cover was cracked on the top.Per functional testing, the device failed the ground bond part of the electrical safety test.The complaint was confirmed.The root cause was a faulty line cord.It was unknown what caused it to become faulty.The product's history records were reviewed and there were no non-conformances nor service-related issues that would have resulted in the reported complaint.No action taken due to the condition of the device.It was deemed beyond economical repair and was scrapped.
 
Event Description
It was reported that the device failed the electrical test.Discovered while doing preventative maintenance (pm) testing.There was no patient involvement and no patient harm/adverse event reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE FLUID WARMER, 390
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan ln n
minneapolis MN 55442
Manufacturer Contact
reed covert
6000 nathan lane north
minneapolis, MN 55442
2247062300
MDR Report Key18934191
MDR Text Key338252206
Report Number3012307300-2024-01458
Device Sequence Number1
Product Code LGZ
UDI-Device Identifier10695085002796
UDI-Public10695085002796
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K911383
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 03/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/19/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHL-390
Device Catalogue NumberCON-HL-390
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/05/2024
Date Manufacturer Received02/21/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/29/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-