• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SISTEMAS MEDICOS ALARIS S.A. DE C.V. BD MAXPLUS NEEDLELESS CONNECTOR; INTRAVASCULAR ADMINISTRATION SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SISTEMAS MEDICOS ALARIS S.A. DE C.V. BD MAXPLUS NEEDLELESS CONNECTOR; INTRAVASCULAR ADMINISTRATION SET Back to Search Results
Catalog Number MP1000 CHINA
Device Problem Complete Blockage (1094)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/30/2023
Event Type  malfunction  
Event Description
It was reported that bd maxplus needleless connector was occluded the following information was received by the initial reporter with the following verbatim: on (b)(6) 2023, the patient suffered from heart failure.He was given infusions as directed by the doctor and established an intravenous indwelling needle to connect to the needleless connector.It was found that the needleless connector was blocked and was replaced immediately.
 
Manufacturer Narrative
H.3.If a device evaluation and/or device history review is completed, a supplemental report will be filed.
 
Manufacturer Narrative
Investigation results: no samples were received for investigation of pr (b)(4), in which the customer has stated: ¿the needleless connector was blocked¿.The product in use at the time is reported to be a mp1000 china from lot 22125382.Further information from the customer stated that the flow was blocked completely and also they confirms that there were no visible damages on the product.The details of this feedback were forwarded to the manufacturing site for investigation, where a review of the production records for lot 22125382 did not identify any in-process testing failures or quality deviations which may have caused or contributed to a report of this nature.The root cause of the customer¿s experience could not be determined in this instance as no sample was received for investigation.Without a sample to examine it is not possible to determine whether a manufacturing defect could have caused or contributed to a report of this nature.Previous investigations have identified that certain designs of male luer can cause flow restriction or occlusion as they can grip onto the piston of the maxplus preventing its proper collapse.This can sometimes be resolved by disconnecting and reattaching the same luer connection which may reposition the luer against the piston and improve the flow, or alternatively by changing the connecting male luer.In this instance, without the complaint sample to examine it has not been possible to conclusively link this feedback to a specific failure mode; however, a review of the customer feedback database indicates that this is a rare occurrence with a small number of similar reports against the mp1000 china products in the past 12 months.
 
Event Description
No additional information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD MAXPLUS NEEDLELESS CONNECTOR
Type of Device
INTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
SISTEMAS MEDICOS ALARIS S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer (Section G)
SISTEMAS MEDICOS ALARIS S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer Contact
helen cox (mdr)
75 north fairway drive
vernon hills, IL 60061
8473935694
MDR Report Key18934655
MDR Text Key338336331
Report Number9616066-2024-00434
Device Sequence Number1
Product Code FPA
UDI-Device Identifier10885403238321
UDI-Public(01)10885403238321
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K072542
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 02/28/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/19/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue NumberMP1000 CHINA
Device Lot Number22125382
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/26/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/12/2022
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-