Catalog Number J-SOS-100500
Device Problems Fluid/Blood Leak (1250); Material Puncture/Hole (1504)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/22/2024
Event Type  malfunction  
Event Description
As reported, the bakri tamponade balloon catheter was tested prior to placement, the balloon had a pin hole leak.The device did not make patient contact.Another device of the same type was used to complete the procedure.A section of the device did not remain inside the patient¿s body.The patient did not require any additional intervention and did not experience any adverse effects due to this occurrence.
 
Manufacturer Narrative
Blank fields on this form indicate the information is unknown or unavailable.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Manufacturer Narrative
Blank fields on this form indicate the information is unknown, unavailable, or unchanged.Investigation ¿ evaluation as reported, the bakri tamponade balloon catheter was tested prior to placement, the balloon had a pin hole leak.The device did not make patient contact.Another device of the same type was used to complete the procedure.A section of the device did not remain inside the patient¿s body.The patient did not require any additional intervention and did not experience any adverse effects due to this occurrence.A document-based investigation was performed including a review of complaint history, device history record (dhr), instructions for use (ifu), and quality control procedures.Additionally, a visual inspection and functional test of the device were conducted.The bakri tamponade balloon catheter was returned for evaluation.A pinhole leak was found in the balloon material.A shadow in the material next to the pinhole was noticed under magnification.A document-based investigation evaluation was performed.A review of the device master record (dmr) concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.A review of the dhr for the reported complaint device lot revealed no recorded non-conformances relevant to the failure mode.A database search did not identify any other complaints associated with the reported device lot.Review of the device history record, complaint history, quality control documents, and device failure analysis does not indicate that the device was manufactured out of specification and does not suggest items in the lot or similar devices in the field or in house are nonconforming.Cook also reviewed product labeling.The product ifu, t_j-sos_rev4 ¿bakri postpartum balloon,¿ provides the following information to the user related to the reported failure mode: how supplied "upon removal from the package, inspect the product to ensure no damage has occurred." based on the information provided, inspection of the returned device, and the results of the investigation, a definitive cause of the event could not be determined.The appropriate personnel have been notified.Cook will continue to monitor for similar complaints.Per the risk assessment no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BAKRI TAMPONADE BALLOON CATHETER
Type of Device
OQY INTRAUTERINE BALLOON
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key18935673
MDR Text Key338081048
Report Number1820334-2024-00376
Device Sequence Number1
Product Code OQY
UDI-Device Identifier10827002306735
UDI-Public(01)10827002306735(17)250616(10)14797247
Combination Product (y/n)N
PMA/PMN Number
K170622
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Distributor
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/19/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberJ-SOS-100500
Device Lot Number14797247
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Date Manufacturer Received03/28/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/16/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient SexFemale
-
-