Catalog Number D134702
Device Problem Contamination /Decontamination Problem (2895)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/16/2023
Event Type  malfunction  
Manufacturer Narrative
The device was returned to biosense webster (bwi) for evaluation.The returned device's visual inspection and microscopic analysis were performed following bwi procedures.Visual analysis revealed a damage in the pebax and traces of blood.Microscopic analysis revealed a hole in the pebax exposing the internal components.The root cause of the damage could be related to improper handling; however, this cannot be conclusively determined.The issue reported by the customer was confirmed.The broken pebax was identified during the investigation of the sample received.A manufacturing record evaluation was performed for the finished device number 30965988l, and no internal action was found during the review.As part of biosense webster's quality process, all devices are manufactured, inspected, and released to approved specifications.This product issue will be addressed through bwi's quality system.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster, inc., or its employees that the report constitutes an admission that the product, biosense webster, inc., or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Manufacturer's reference number: (b)(4).
 
Event Description
It was reported that a patient underwent a ventricular tachycardia (vt) cardiac ablation procedure with a thermocool® smart touch® sf uni-directional navigation catheter for which the biosense webster¿s product analysis lab (pal) observed a hole in the pebax exposing the internal components.Initially, it was reported that blood in catheter spring was observed during the procedure.A different catheter was used, and the issue was resolved.No adverse patient consequence was reported.The issue with blood found in the catheter was assessed as non mdr reportable.The potential risk that it could cause or contribute to a serious injury or death to the operator or patient was remote.The biosense webster, inc.Product analysis lab received the device for evaluation and per the evaluation completion on (b)(6) 2024, there was a hole in the pebax exposing the internal components.This was assessed as mdr reportable with an awareness date of 21-feb-2024.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THERMOCOOL® SMART TOUCH® SF UNI-DIRECTIONAL NAVIGATION CATHETER
Type of Device
CARDIAC ABLATION PERCUTANEOUS CATHETER
Manufacturer (Section D)
BIOSENSE WEBSTER INC
31 technology drive, suite 200
irvine CA 92618
Manufacturer (Section G)
BIOSENSE WEBSTER INC (JUAREZ)
circuito interior norte
1820parque industrial salvacar
juarez 32599
MX   32599
Manufacturer Contact
kate karberg
31 technology dr
irvine, CA 92618
3035526892
MDR Report Key18935761
MDR Text Key338558280
Report Number2029046-2024-00921
Device Sequence Number1
Product Code LPB
UDI-Device Identifier10846835009781
UDI-Public10846835009781
Combination Product (y/n)N
Reporter Country CodeNL
PMA/PMN Number
P030031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other
Type of Report Initial
Report Date 03/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/19/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberD134702
Device Lot Number30965988L
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/07/2024
Date Manufacturer Received02/21/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/12/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial