• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTEGRA LIFESCIENCES (IRELAND) LTD. MLX 300W XENON LIGHTSOURCE; N/A

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTEGRA LIFESCIENCES (IRELAND) LTD. MLX 300W XENON LIGHTSOURCE; N/A Back to Search Results
Catalog Number 00MLX
Device Problems Overheating of Device (1437); Separation Problem (4043)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The mlx 300w xenon lightsource (00mlx) was returned for evaluation: failure analysis - the mlx 300w xenon lightsource was received in used condition.Evaluation of the xenon lightsource identified that the unit failed the turret rotation, the cable insertion test, the ac ferrite inside the unit was damaged.In addition, the bezel had internal damage.The turret, ac ferrite and the bezel have been replaced to correct these issues.Root cause analysis.The reported complaint was confirmed.It was determined that the issue of the light cord selector being burnt could be due to damage caused by rough handling/environmental damage.No manufacturing, workmanship, or material deficiency has been identified.
 
Event Description
A facility reported that the light cord selector of the mlx 300w xenon lightsource (00mlx) was burnt.In addition, the light filter had fallen out of its holder.It was noted that there was no damage to the filter or holder.It is unknown under what circumstance this event was discovered; however, no patient injury, death or surgical delay was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MLX 300W XENON LIGHTSOURCE
Type of Device
N/A
Manufacturer (Section D)
INTEGRA LIFESCIENCES (IRELAND) LTD.
sragh tullamore, co. offaly
EI 
Manufacturer (Section G)
INTEGRA LIFESCIENCES (IRELAND) LTD.
sragh tullamore, co. offaly
EI  
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key18937227
MDR Text Key338451562
Report Number3006697299-2024-00028
Device Sequence Number1
Product Code FCW
UDI-Device Identifier10381780168140
UDI-Public10381780168140
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 03/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/19/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number00MLX
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/02/2024
Date Manufacturer Received02/22/2024
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-