The mlx 300w xenon lightsource (00mlx) was returned for evaluation: failure analysis - the mlx 300w xenon lightsource was received in used condition.Evaluation of the xenon lightsource identified that the unit failed the turret rotation, the cable insertion test, the ac ferrite inside the unit was damaged.In addition, the bezel had internal damage.The turret, ac ferrite and the bezel have been replaced to correct these issues.Root cause analysis.The reported complaint was confirmed.It was determined that the issue of the light cord selector being burnt could be due to damage caused by rough handling/environmental damage.No manufacturing, workmanship, or material deficiency has been identified.
|
A facility reported that the light cord selector of the mlx 300w xenon lightsource (00mlx) was burnt.In addition, the light filter had fallen out of its holder.It was noted that there was no damage to the filter or holder.It is unknown under what circumstance this event was discovered; however, no patient injury, death or surgical delay was reported.
|