It was observed that during the device evaluation, the duodenovideoscope aw-cylinder (air, water) had foreign objects.The aw-tube (air, water) had foreign objects.The distal end had foreign objects and was cracked.There were no reports of patient harm.
|
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.Additional information added to fields h3 and h6.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 19 years since the subject device was manufactured.The legal manufacture was unable to confirm whether the customer's reprocessing steps were deviated from the instructions for use.Based on the results of the investigation, olympus confirmed foreign material remained in device, but the type of the material cannot be identified.For the events of foreign material in air/water cylinder u, air/water channel and distal end, we presume from the following investigation results that reprocessing was not conducted properly, subsequently, the material was not possibly eliminated.The following nonconformity was confirmed from (sales) service business center inspection results: (air leak inspection) fail: due to damage on switch 1, water tightness is lost.Also, based on the results of the investigation, it is likely that irregular stress was applied to the distal end during device handling by the user.Subsequently, the suggested event of distal end cracked may have occurred.The following was written in instructions for use to warn improper reprocessing in regards of the events of foreign material in air/water cylinder u, foreign material in air/water channel and foreign material at distal end: "an insufficiently reprocessed endoscope and/or accessory may pose an infection control risk to the patients and/or operators who touch them." the even of distal end cracked can be detected/prevented by following the instructions for use: instructions for use: evis exera tjf type 145 operation manual _important information ¿ please read before use_ warnings and cautions :do not strike, bend, hit, pull, twist, or drop the endoscope¿s distal end, insertion tube, bending section, control section, universal cord, or endoscope connector of the endoscope with excessive force.The endoscope may be damaged and could cause patient injury, burns, bleeding and/or perforations.It could also cause parts of the endoscope to fall off inside the patient.Three attempts were performed to obtain additional information, but no response was received from the customer.Olympus will continue to monitor field performance for this device.
|