AIZU OLYMPUS CO., LTD. EVIS EXERA DUODENOVIDEOSCOPE Back to Search Results
Model Number TJF-145
Device Problems Crack (1135); Failure to Clean Adequately (4048)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The device evaluation is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was observed that during the device evaluation, the duodenovideoscope aw-cylinder (air, water) had foreign objects.The aw-tube (air, water) had foreign objects.The distal end had foreign objects and was cracked.There were no reports of patient harm.
 
Manufacturer Narrative
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.Additional information added to fields h3 and h6.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 19 years since the subject device was manufactured.The legal manufacture was unable to confirm whether the customer's reprocessing steps were deviated from the instructions for use.Based on the results of the investigation, olympus confirmed foreign material remained in device, but the type of the material cannot be identified.For the events of foreign material in air/water cylinder u, air/water channel and distal end, we presume from the following investigation results that reprocessing was not conducted properly, subsequently, the material was not possibly eliminated.The following nonconformity was confirmed from (sales) service business center inspection results: (air leak inspection) fail: due to damage on switch 1, water tightness is lost.Also, based on the results of the investigation, it is likely that irregular stress was applied to the distal end during device handling by the user.Subsequently, the suggested event of distal end cracked may have occurred.The following was written in instructions for use to warn improper reprocessing in regards of the events of foreign material in air/water cylinder u, foreign material in air/water channel and foreign material at distal end: "an insufficiently reprocessed endoscope and/or accessory may pose an infection control risk to the patients and/or operators who touch them." the even of distal end cracked can be detected/prevented by following the instructions for use: instructions for use: evis exera tjf type 145 operation manual _important information ¿ please read before use_ warnings and cautions :do not strike, bend, hit, pull, twist, or drop the endoscope¿s distal end, insertion tube, bending section, control section, universal cord, or endoscope connector of the endoscope with excessive force.The endoscope may be damaged and could cause patient injury, burns, bleeding and/or perforations.It could also cause parts of the endoscope to fall off inside the patient.Three attempts were performed to obtain additional information, but no response was received from the customer.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA DUODENOVIDEOSCOPE
Type of Device
DUODENOVIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18938226
MDR Text Key338186156
Report Number9610595-2024-05876
Device Sequence Number1
Product Code FDT
UDI-Device Identifier04953170339950
UDI-Public04953170339950
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
K954451
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 04/24/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/19/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberTJF-145
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/05/2024
Was the Report Sent to FDA? No
Date Manufacturer Received04/19/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/14/2004
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-