Model Number 7842
Device Problems Failure to Capture (1081); Mechanical Problem (1384); Off-Label Use (1494); Difficult to Remove (1528); Device Dislodged or Dislocated (2923); Material Integrity Problem (2978); High Capture Threshold (3266)
Patient Problem Cardiac Perforation (2513)
Event Date 02/15/2024
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that this lead implanted in an off-label way at the left bundle branch (lbb) found intermittently capturing at maximum output.An x-ray analysis showed lbb lead may have moved forward.A new lead was implanted in the lbb, nonetheless, after testing of the newly implanted lead, the old lbb lead was attempted to be removed.Over 60 turns of the helix, body turns and manipulation with firm stylet were done but the helix did not retract.X-ray showed the helix stretched into a spike or hook.Doctor decided to abandon lead and plug it into the inactive right atrial lead port.The lead was turned off to allow future magnetic resonance imaging.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INGEVITY+
Type of Device
IMPLANTABLE LEAD
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
road 698, lot no. 12
dorado PR 00646 -260
*   00646-2602
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key18938759
MDR Text Key338066796
Report Number2124215-2024-16799
Device Sequence Number1
Product Code NVN
UDI-Device Identifier00802526604522
UDI-Public00802526604522
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 03/20/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/20/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number7842
Device Catalogue Number7842
Device Lot Number1267677
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/28/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/07/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Life Threatening; Hospitalization;
Patient Age72 YR
Patient SexMale
-
-