BOSTON SCIENTIFIC CORPORATION VALITUDE X4 CRT-P; PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P)
|
Back to Search Results |
|
Model Number U128 |
Device Problems
High impedance (1291); Over-Sensing (1438); Pacing Problem (1439)
|
Patient Problem
Syncope/Fainting (4411)
|
Event Date 01/30/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
If pertinent information is provided in the future, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that this device and right ventricular (rv) lead exhibited pace impedance measurements of greater than 3,000 ohms, resulting in a lead safety switch (lss).This patient experienced a ventricular episode.Technical services (ts) reviewed noting there appeared to be myopotential oversensing occurring.This patient passed out during this episode.A previous signal artifact monitor (sam) episode was observed.The health care professional (hcp) reprogrammed the sensitivity as this patient is pacemaker dependent.This device remains in service.No adverse patient effects were reported.
|
|