As reported by an edwards lifesciences field representative, during a right-transfemoral transcatheter aortic valve replacement procedure with a 26mm sapien 3 ultra valve, the 26mm commander delivery system balloon ruptured at max inflation using nominal volume.The perceived root cause of the balloon burst was calcification at the sinotubular junction.During delivery system withdrawal, the balloon would not retract into the 14fr esheath+.Preexisting contralateral access was utilized to snare the balloon tip, allowing for retraction of the delivery system into the sheath.Cutdown was performed to surgically close the vessel as "the sheath end was much larger than standard case" and the operator did not want a bleeding complication.The sheath and delivery system were removed as a unit.Upon removal, it was noted that the sheath had full delamination.It was also noticed that a portion of the delivery system balloon remained in the body.The balloon material was found at the access site and was ultimately retrieved from the arteriotomy and surgical closure with a patch was successful.All balloon material was accounted for and successfully removed from the patient.The patient was discharged without issue.
|