• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MPRI CAPSUREFIX NOVUS MRI SURESCAN; PERMANENT PACEMAKER ELECTRODE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MPRI CAPSUREFIX NOVUS MRI SURESCAN; PERMANENT PACEMAKER ELECTRODE Back to Search Results
Model Number 407652
Device Problems Adverse Event Without Identified Device or Use Problem (2993); High Capture Threshold (3266)
Patient Problems Cardiac Tamponade (2226); Cardiac Perforation (2513)
Event Date 03/06/2024
Event Type  Injury  
Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event Description
It was reported one week post implant that the patient experienced cardiac tamponade.It was found that the right atrial (ra) lead exhibited a cardiac perforation and high thresholds.The ra lead was reprogrammed and later was explanted and replaced. pericardiocentesis was performed. no further patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product event summary: the full lead was returned and analyzed.Returned product analysis was performed and no anomalies were found.Visual analysis of the lead indicated apparent explant damage.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPSUREFIX NOVUS MRI SURESCAN
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer (Section G)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key18941288
MDR Text Key338136212
Report Number2649622-2024-07679
Device Sequence Number1
Product Code DTB
UDI-Device Identifier00643169708198
UDI-Public00643169708198
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P930039
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/20/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number407652
Device Catalogue Number407652
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/08/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/25/2024
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
6935M62 LEAD, DTPA2QQ CRTD, 429888 LEAD
Patient Outcome(s) Hospitalization; Life Threatening; Required Intervention;
Patient Age62 YR
Patient SexMale
-
-