• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS NORTH AMERICA LLC HEARTSTART XL+ DEFIBRILLATOR/MONITOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS NORTH AMERICA LLC HEARTSTART XL+ DEFIBRILLATOR/MONITOR Back to Search Results
Model Number 861290
Device Problem Charging Problem (2892)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/26/2024
Event Type  malfunction  
Event Description
Philips received a complaint on the xl+ device indicating that there was a charge fault.There was no patient involvement.The fse evaluated the device on site.The error message of charge fault was observed.The fse plugged and unplugged the battery, restarted the device and ran the self-test and all checks passed.Although the issue could not be replicated, this will be documented as a malfunction that occurred at the time of the reported event, the cause of which was not determined.The device remains at the customer site and no further evaluation is warranted at this time.Based on the information available and the testing conducted, the cause of the reported problem was not confirmed, could not duplicate reported issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTSTART XL+ DEFIBRILLATOR/MONITOR
Type of Device
DEFIBRILLATOR
Manufacturer (Section D)
PHILIPS NORTH AMERICA LLC
22100 bothell everett highway
bothell WA 98021
Manufacturer (Section G)
PHILIPS NORTH AMERICA LLC
22100 bothell everett highway
bothell WA 98021
Manufacturer Contact
tanya deschmidt
22100 bothell everett highway
bothell, WA 98021
9095703538
MDR Report Key18942226
MDR Text Key338261192
Report Number3030677-2024-01050
Device Sequence Number1
Product Code MKJ
UDI-Device Identifier00884838023680
UDI-Public00884838023680
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K110825
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/20/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number861290
Device Catalogue Number861290
Was Device Available for Evaluation? Yes
Date Manufacturer Received02/26/2024
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-