Model Number 6172
Device Problem Low impedance (2285)
Patient Problems Inadequate Pain Relief (2388); Movement Disorder (4412)
Event Date 01/12/2024
Event Type  Injury  
Event Description
Related manufacturer reference number: 3006705815-2024-02138.It was reported that patient experienced a fluid intrusion at the ipg site, patient also experienced ineffective therapy and was unable to enter mri mode.Patients lead had low impedances.As a result, surgical intervention was undertaken on (b)(6) 2024 wherein ipg and lead were explanted and replaced to address the issue.
 
Manufacturer Narrative
Date of event is estimated.
 
Event Description
Additional information indicates that lead was not explanted and replaced lead was disconnected and whipped off to address the issue.
 
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
8CH INFINITY DBS LEAD KIT, 40CM, 0.5, B
Type of Device
DBS LEAD
Manufacturer (Section D)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
Manufacturer Contact
heidi syndergaard
6901 preston road
plano, TX 75024
9723098000
MDR Report Key18944458
MDR Text Key338185369
Report Number1627487-2024-07616
Device Sequence Number1
Product Code MHY
UDI-Device Identifier05415067030320
UDI-Public05415067030320
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P140009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/20/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date11/28/2020
Device Model Number6172
Device Lot Number6742505
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/25/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/29/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
DBS BURR HOLE CAP; DBS EXTENSION
Patient Outcome(s) Other;
Patient SexMale
Patient Weight98 KG
-
-