• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SHIRAKAWA OLYMPUS CO., LTD. EVIS EXERA II ULTRASOUND GASTROVIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SHIRAKAWA OLYMPUS CO., LTD. EVIS EXERA II ULTRASOUND GASTROVIDEOSCOPE Back to Search Results
Model Number GF-UCT180
Device Problem Failure to Clean Adequately (4048)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The device evaluation is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was observed that during the device inspection, the ultrasound gastrovideoscope exhibited a clogged nozzle blocking air/water flow.There were no reports of patient involvement.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the foreign material could not be identified nor the root cause of it.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA II ULTRASOUND GASTROVIDEOSCOPE
Type of Device
ULTRASOUND GASTROVIDEOSCOPE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18945888
MDR Text Key338684294
Report Number3002808148-2024-02638
Device Sequence Number1
Product Code ODG
UDI-Device Identifier04953170356339
UDI-Public04953170356339
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K093395
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 04/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/20/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberGF-UCT180
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/12/2024
Was the Report Sent to FDA? No
Date Manufacturer Received04/05/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured01/11/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-