BAXTER HEALTHCARE CORPORATION AMIA AUTOMATED PD CYCLER SET; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
|
Back to Search Results |
|
Catalog Number 5C5479 |
Device Problems
Material Puncture/Hole (1504); Protective Measures Problem (3015)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/20/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
An alarm indicative of a potential malfunction of the disposable cassette was reported.The device was not returned and the lot number is unknown; therefore, a device analysis could not be completed.However, a hole was reported in an unspecified line of the amia cassette, which is known to cause this alarm.The cause of the hole could not be determined.Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that an amia device experienced an unspecified alarm; further described as "error in line".It was further reported that "the patient suspected a hole or air in line, however uncertain".The patient was connected at the time of the alarm.This occurred during an unspecified process step of peritoneal dialysis therapy.The patient disconnected and ended the therapy session.There was no report of patient injury, however, the patient was given a prophylactic antibiotic (one-time dose of intraperitoneal ancef, one gram) as precautionary measure.Additionally, the patient¿s transfer set was changed as a precautionary measure.No additional information is available.
|
|