• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION AMIA AUTOMATED PD CYCLER SET; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION AMIA AUTOMATED PD CYCLER SET; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY Back to Search Results
Catalog Number 5C5479
Device Problems Material Puncture/Hole (1504); Protective Measures Problem (3015)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/20/2024
Event Type  malfunction  
Manufacturer Narrative
An alarm indicative of a potential malfunction of the disposable cassette was reported.The device was not returned and the lot number is unknown; therefore, a device analysis could not be completed.However, a hole was reported in an unspecified line of the amia cassette, which is known to cause this alarm.The cause of the hole could not be determined.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that an amia device experienced an unspecified alarm; further described as "error in line".It was further reported that "the patient suspected a hole or air in line, however uncertain".The patient was connected at the time of the alarm.This occurred during an unspecified process step of peritoneal dialysis therapy.The patient disconnected and ended the therapy session.There was no report of patient injury, however, the patient was given a prophylactic antibiotic (one-time dose of intraperitoneal ancef, one gram) as precautionary measure.Additionally, the patient¿s transfer set was changed as a precautionary measure.No additional information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMIA AUTOMATED PD CYCLER SET
Type of Device
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - MOUNTAIN HOME
1900 n highway 201
mountain home AR 72653
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key18946360
MDR Text Key338220070
Report Number1416980-2024-01247
Device Sequence Number1
Product Code FKX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K151525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Nurse
Type of Report Initial
Report Date 03/20/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/20/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number5C5479
Device Lot NumberASKU
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/20/2024
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
AMIA AUTOMATED PD SYSTEM
-
-