ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V. 32" (81 CM) APPX 6.3 ML, 15 DROP ADMIN SET W/0.2 MICRON FILTER, ROTATING LUER W/; STOPCOCK, I.V. SET
|
Back to Search Results |
|
Catalog Number B30183 |
Device Problem
Air/Gas in Device (4062)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/14/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
One (1) used sample list #b30183 connected with a used, equashield, and used, bag spike adaptor; lot #unknown and one used.A 0.9% sodium chloride 500ml bag was returned for evaluation.As received the inlet filter appeared to be wetted out.No additional damage or anomalies were observed.The set was primed and purged as per procedure and a leak coming from the inlet filter vent was confirmed.However, no air beyond the outlet filter vent to the patient line was confirmed.Complaints of product incorporates/allows air passage cannot be confirmed.However.The probable cause of the leaks coming from the inlet filter is typically a temporary or complete loss of hydrophobic properties of the filter vent material due to an infusate interaction during use.The lot history was reviewed and no nonconformities were identified that may have contributed to the reported complaint.
|
|
Event Description
|
It was reported that a 32" (81 cm) appx 6.3 ml, 15 drop admin set w/0.2 micron filter, rotating luer w/filter cap, bag hanger allowed air to enter the tubing, trapping fluid above in the bag and drip chamber during patient use.Then air in the tubing down to the filter and there is fluid below the filter.They are unable to back prime to clear the line of air and run the medicine.The device is contaminated with chemotherapy.There was a patient involved, no patient harm and there was a delay in therapy as the drug had to be remade to complete the therapy.No additional information is available at this time.
|
|
Search Alerts/Recalls
|
|
|