• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE MOLECULAR SYSTEMS, INC. - BRANCHBURG COBAS® SARS-COV-2 & INFLUENZA A/B; COVID-19 MULTI-ANALYTE RESPIRATORY PANEL NUCLEIC ACID DEVICES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE MOLECULAR SYSTEMS, INC. - BRANCHBURG COBAS® SARS-COV-2 & INFLUENZA A/B; COVID-19 MULTI-ANALYTE RESPIRATORY PANEL NUCLEIC ACID DEVICES Back to Search Results
Catalog Number 09211101190
Device Problem Non Reproducible Results (4029)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/03/2024
Event Type  malfunction  
Event Description
In light of the covid-19 pandemic and the subsequent emergency use authorizations (euas) for sars-cov-2 diagnostic tests, the agency has requested heightened reporting beyond the reasonably suggests requirements of 803 to include allegations of false positive or false negative results independent of harm or malfunction or off-label use.Pursuant to the agency¿s instruction, we hereby submit this mdr.A customer alleged a discrepant result for six patient samples while using the cobas sars-cov-2 & influenza a/b nucleic acid test on the cobas liat system.The alleged samples initially generated a positive result for sars-cov-2.Fresh samples from the patients were tested on the same cobas liat analyzer which yielded a negative result for all targets.The initial positive results were not released.No harm was alleged.An investigation was conducted to evaluate the customer issue.Per fda¿s eua guidance, 6 mdrs will be filed.
 
Manufacturer Narrative
The liat analyzer serial number was (b)(6).Although a root cause cannot be definitively identified, the data showed true amplification for the positive results.A result discrepancy can occur when a sample generates a ct value at the limit of detection (lod) of the assay.Additionally, it is possible that the observed discrepancies that did not generate a ct value at the limit of detection are due to workflow contamination.This is a run/workflow-specific issue and the reagent is performing as expected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COBAS® SARS-COV-2 & INFLUENZA A/B
Type of Device
COVID-19 MULTI-ANALYTE RESPIRATORY PANEL NUCLEIC ACID DEVICES
Manufacturer (Section D)
ROCHE MOLECULAR SYSTEMS, INC. - BRANCHBURG
1080 us highway 202 south
branchburg NJ 08876
Manufacturer (Section G)
ROCHE MOLECULAR SYSTEMS, INC. - BRANCHBURG
1080 us highway 202 south
branchburg NJ 08876
Manufacturer Contact
timothy blair
1080 us hwy 202 s
branchburg, NJ 08876
MDR Report Key18948000
MDR Text Key338337100
Report Number2243471-2024-00834
Device Sequence Number1
Product Code QLT
Combination Product (y/n)Y
Reporter Country CodeJA
PMA/PMN Number
EUA201779
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/21/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/21/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number09211101190
Device Lot Number30307Z
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/05/2024
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-