Model Number WR9200 |
Device Problems
Display or Visual Feedback Problem (1184); Failure to Power Up (1476); Operating System Becomes Nonfunctional (2996)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/14/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
H3.Analysis of the wireless recharger (s/n (b)(6)) found database corruption during out of box setup resulting in the device running in runmode.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Event Description
|
It was reported that manufacturer representative (rep) was setting up for a case today and attempted to prep the wireless recharger.They plugged into the wall and set the recharger on the dock but there were no lights.The rep removed and placed it back on the dock, but there were no lights.The rep then attempted to turn on the wr, but there was response.The rep used a second wr which worked fine.
|
|