Last week and this morning, my technicians came to me to report defective tablo cartridges.The leaks both appear to be coming from the connection from the hard plastic coupler that attaches to the soft tubing that wraps around the blood pump, to the rest of the tubing.This is the same area as the blood leaking from the cartridge that was reported earlier this year.The lot number in that case was m23137l02s01.On the second instance, the lot number was m23156l03s01.Manufacturer response for set, tubing, blood, with and without anti-regurgitation valve, tablo(r) cartridge (per site reporter).No response yet.
|