It was reported that the cardiosave intra-aortic balloon pump (iabp) had recorder interface board related malfunction.The patient involvement is unknown.Getinge field service engineer (fse) investigated the customer's complaint printer would only print a few inches then stop, replaced printe, tether assembly was fully extended and would not retract, replaced assembly.Pneumatic module assy leak found during testing.Replaced pneumatic module assy.Unit passed all calibration, functional and safety tests performed.Unit was returned to customer and cleared for clinical use.Demand pm done at the same time.See that so for fall checkout information.
|