• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC NEUROMODULATION PRECISION MONTAGE MRI; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC NEUROMODULATION PRECISION MONTAGE MRI; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF Back to Search Results
Model Number SC-1200
Device Problems Use of Device Problem (1670); Failure to Disconnect (2541)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/23/2024
Event Type  malfunction  
Event Description
It was reported that during a lead revision procedure, (see mfr.Report 3006630150-2024-01644) the set screw of the implantable pulse generator (ipg) was backed out and the lead was unable to be released from the ipg.It was noted that the avista lead was pulled out of the implantable pulse generator (ipg) header with no problem.After the set screw was opened too far, the screw came out.Even when the screw was not in the header anymore, and the physician pulled very hard, it was not possible to pull out the linear lead.This resulted in the lead being demolished, and therefore neither the lead nor the ipg could be used any longer.In addition, the avista lead had to be removed and exchanged as it was not possible to insert a stylet into the avista lead in order to push it back in place.Therefore, a new ipg and two new linear leads were implanted.The patient was doing fine postoperatively.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRECISION MONTAGE MRI
Type of Device
STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF
Manufacturer (Section D)
BOSTON SCIENTIFIC NEUROMODULATION
25155 rye canyon loop
valencia CA 91355
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
erik shernburne
25155 rye canyon loop
valencia, CA 91355
7632920920
MDR Report Key18948721
MDR Text Key339151639
Report Number3006630150-2024-01667
Device Sequence Number1
Product Code LGW
UDI-Device Identifier08714729905943
UDI-Public08714729905943
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P030017
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 03/21/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/21/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date05/05/2022
Device Model NumberSC-1200
Device Catalogue NumberSC-1200
Device Lot Number369332
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/08/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/05/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-