B. BRAUN MELSUNGEN AG EASYPUMP; PUMP, INFUSION, ELASTOMERI Back to Search Results
Catalog Number 4540018-02
Device Problem Excess Flow or Over-Infusion (1311)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/14/2024
Event Type  malfunction  
Event Description
As reported by the user facility: event 3.Brief inquiry description: easypump infusing faster than intended.Detailed inquiry description: was contacted by director of pharmacy.She said that they have had at least 7 infusions with easypump finish within a 24 hr window (about twice as fast as intended) over the last 2 weeks.The total amount of easypump therapies delivered in this time period was 24.We reviewed preparation protocols and patient education.No injury reported.
 
Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).The investigation is ongoing at this time.A follow-up will be submitted when the investigation results become available.
 
Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).A device history record (dhr) review was performed for the reported lot number and no abnormalities or non-conformances were noted during the in process or final product inspection.No sample was provided for evaluation.Based on the data from the investigation we are unable to determine the root cause of the reported incident.The reported defect was unable to be confirmed.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EASYPUMP
Type of Device
PUMP, INFUSION, ELASTOMERI
Manufacturer (Section D)
B. BRAUN MELSUNGEN AG
carl-braun strabe 1
melsungen, 34212
GM  34212
Manufacturer (Section G)
B. BRAUN MELSUNGEN AG
carl-braun strabe 1
melsungen, 34212
GM   34212
Manufacturer Contact
jonathan severino
901 marcon blvd
allentown, PA 18109
4847197287
MDR Report Key18949030
MDR Text Key338248910
Report Number9610825-2024-00205
Device Sequence Number1
Product Code MEB
UDI-Device Identifier04046964448638
UDI-Public(01)04046964448638
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K081905
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/21/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/21/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number4540018-02
Device Lot Number23K10GE561
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/15/2024
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-