• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDIANA CO. LTD. NELLCOR; OXIMETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDIANA CO. LTD. NELLCOR; OXIMETER Back to Search Results
Model Number 10005941
Device Problems Failure to Power Up (1476); Charging Problem (2892); Temperature Problem (3022)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/06/2023
Event Type  malfunction  
Event Description
According to the reporter, the product had a charging failure and it turned itself off prior to use. there was no patient involvement.Medtronic's initial evaluation of the incident device found there was a burnt trace on the circuit board.
 
Manufacturer Narrative
H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the printed circuit board was assembled incorrectly.Additionally, the battery was defective, there was a burnt trace on the circuit board and there was damage to the device but it did not interfere with the unit functioning correctly.It was reported that the monitor did not charge the battery and the unit powered down unexpectedly but then turned back on by itself.The reported issues were confirmed.The most likely cause was traced to a component failure.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NELLCOR
Type of Device
OXIMETER
Manufacturer (Section D)
MEDIANA CO. LTD.
1650 1 2 donghwa ri munmak eu
kangwon do wonju si 00000
KS  00000
Manufacturer (Section G)
MEDIANA CO. LTD.
1650 1 2 donghwa ri munmak eu
kangwon do wonju si 00000
KS   00000
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key18949072
MDR Text Key338239945
Report Number2936999-2024-00488
Device Sequence Number1
Product Code DQA
UDI-Device Identifier10884521163454
UDI-Public10884521163454
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
K120773
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign
Reporter Occupation Other
Type of Report Initial
Report Date 03/21/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/21/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10005941
Device Catalogue Number10005941
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/05/2024
Date Manufacturer Received03/12/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/03/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-