Model Number U228 |
Device Problems
Signal Artifact/Noise (1036); High impedance (1291); Over-Sensing (1438); Pacing Problem (1439)
|
Patient Problem
Presyncope (4410)
|
Event Date 11/01/2023 |
Event Type
Injury
|
Event Description
|
It was reported that this cardiac resynchronization therapy pacemaker (crt-p) system triggered a lead safety switch (lss) due to high out of range right ventricular (rv) impedance measurements greater than 2000 ohms, causing the device to switch from bipolar to unipolar configuration, which resulted in oversensing, pacing inhibition and presyncope for the patient.Boston scientific technical services (ts) reviewed data from the device and found noisy signals likely due to unipolar configuration, and also indicated that for the past 12 months, the rv pacing impedance trend includes high impedance peaks.Upon further troubleshooting and device interrogation, no noise or abnormalities were observed when in bipolar sensing.Of note, the implanted rv lead is a non-boston scientific product, and ts has two hypotheses for the reported observations; the first one being lead damage and the second one is a fretting problem between the lead and the device header.Further troubleshooting steps were provided in order to evaluate system integrity and if abnormalities are found, ts recommended lead replacement.At this time, no further information is available.Evidence suggests that this patient was hospitalized as they are pacing dependent.The device remains in service and no additional adverse patient effects have been reported.
|
|
Manufacturer Narrative
|
This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.The associated investigation determined that this device exhibited intermittent impedance measurements with no conclusive evidence of a product performance issue or inadequate lead-to-device connection; please refer to the description for more information regarding the specific circumstances of this event.Investigation has determined that this type of event is likely the result of an intermittent high impedance condition associated with the device spring contact and lead terminal ring.A design enhancement was implemented in 2020 to stabilize the electrical connection between the spring contact and the lead terminal.
|
|
Event Description
|
It was reported that this cardiac resynchronization therapy pacemaker (crt-p) system triggered a lead safety switch (lss) due to high out of range right ventricular (rv) impedance measurements greater than 2000 ohms, causing the device to switch from bipolar to unipolar configuration, which resulted in oversensing, pacing inhibition and presyncope for the patient.Boston scientific technical services (ts) reviewed data from the device and found noisy signals likely due to unipolar configuration, and also indicated that for the past 12 months, the rv pacing impedance trend includes high impedance peaks.Upon further troubleshooting and device interrogation, no noise or abnormalities were observed when in bipolar sensing.Of note, the implanted rv lead is a non-boston scientific product, and ts has two hypotheses for the reported observations; the first one being lead damage and the second one is a fretting problem between the lead and the device header.Further troubleshooting steps were provided in order to evaluate system integrity and if abnormalities are found, ts recommended lead replacement.At this time, no further information is available.Evidence suggests that this patient was hospitalized as they are pacing dependent.The device remains in service and no additional adverse patient effects have been reported.
|
|
Search Alerts/Recalls
|
|