Model Number P1900D004555 |
Device Problem
Activation Problem (4042)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/23/2024 |
Event Type
malfunction
|
Event Description
|
Baxter received a report from a baxter technician stating the cpr was not working and could not be activated.The bed was located at the account.There was no patient user injury reported.This report was filed in our complaint handling system as complaint #: (b)(4).
|
|
Manufacturer Narrative
|
The baxter technician found the cpr valve needed to be replaced.Per the hillrom service manual the totalcare bariatric bed and totalcare bed system requires an effective maintenance program.We recommend that you perform a semi-annual preventative maintenance.Test the cpr release for correct operation and reset of the head cylinder.A search of the baxter maintenance records did not show baxter performed any preventative maintenance on this bed.It is unknown if the facility performs preventative maintenance on their beds.The technician replaced the cpr valve to resolve the reported event.Based on this information, no further action is required.
|
|