U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HILL-ROM BATESVILLE TOTAL CARE FRAME; BED, AC-POWERED ADJUSTABLE HOSPITAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HILL-ROM BATESVILLE TOTAL CARE FRAME; BED, AC-POWERED ADJUSTABLE HOSPITAL Back to Search Results
Model Number P1900D004555
Device Problem Activation Problem (4042)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/23/2024
Event Type  malfunction  
Event Description
Baxter received a report from a baxter technician stating the cpr was not working and could not be activated.The bed was located at the account.There was no patient user injury reported.This report was filed in our complaint handling system as complaint #: (b)(4).
 
Manufacturer Narrative
The baxter technician found the cpr valve needed to be replaced.Per the hillrom service manual the totalcare bariatric bed and totalcare bed system requires an effective maintenance program.We recommend that you perform a semi-annual preventative maintenance.Test the cpr release for correct operation and reset of the head cylinder.A search of the baxter maintenance records did not show baxter performed any preventative maintenance on this bed.It is unknown if the facility performs preventative maintenance on their beds.The technician replaced the cpr valve to resolve the reported event.Based on this information, no further action is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TOTAL CARE FRAME
Type of Device
BED, AC-POWERED ADJUSTABLE HOSPITAL
Manufacturer (Section D)
HILL-ROM BATESVILLE
1069 state route 46 east
batesville IN 47006
Manufacturer Contact
brad wheeler
1069 state route 46 east
batesville, IN 47006
3128199307
MDR Report Key18949756
MDR Text Key338359703
Report Number1824206-2024-00375
Device Sequence Number1
Product Code FNL
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K962942
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 03/21/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/21/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberP1900D004555
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/23/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/10/2002
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse