The event involved a 43 cm (17") transfer set, pur w/clave¿, 10 units where it was reported the infusion took longer than the desired infusion time.During the infusion of paclitaxel, the nurses noticed decreased flow due to an "air occlusion" alarm was triggered, signaling the presences of air in the tubing.The line was primed to expel the air and infuse at the desired flow rate.The patient's condition was unchanged before and after the infusion and no medical intervention needed.There was no delay in therapy, just an extension of the infusion time, therefore a longer consultation for the patient.There was air present in the tubing, which afterwards was removed during priming.It is unknown if the reduction in flow was linked to the air in the tubing or after the pump has alarmed.There was patient involvement, however, no report of patient harm.
|