• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTEGRA LIFESCIENCES MANSFIELD CODMAN DISPOS PERFORATOR; DISPOSABLE PERFORATORS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTEGRA LIFESCIENCES MANSFIELD CODMAN DISPOS PERFORATOR; DISPOSABLE PERFORATORS Back to Search Results
Catalog Number 261221
Device Problem Fail-Safe Did Not Operate (4046)
Patient Problem Brain Injury (2219)
Event Date 02/29/2024
Event Type  Injury  
Manufacturer Narrative
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
 
Event Description
A facility reported: "the perforator did not self-blocked and broke the dura entering the brain".Surgery was delayed for about 5 minutes.Details about injury: cerebral contusion and dural/cerebral laceration.The patient's underlying medical condition is: "coma".The drill used with the perforator is an electric medtronic.The perforator clicked in place in the drill, the recommended spring tests were performed between each burr hole.The angle of approach was perpendicular (as instructions for use states).A perpendicular approach was maintained through the drilling process and there was a constant downward pressure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CODMAN DISPOS PERFORATOR
Type of Device
DISPOSABLE PERFORATORS
Manufacturer (Section D)
INTEGRA LIFESCIENCES MANSFIELD
11 cabot boulevard
11 cabot boulevard
mansfield MA
Manufacturer (Section G)
INTEGRA LIFESCIENCES MANSFIELD
11 cabot boulevard
mansfield MA
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key18950852
MDR Text Key338305895
Report Number3014334038-2024-00058
Device Sequence Number1
Product Code HBF
Combination Product (y/n)N
PMA/PMN Number
K791101
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 03/21/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/21/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number261221
Device Lot Number6899412
Date Manufacturer Received02/29/2024
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-