PROCEPT BIOROBOTICS CORPORATION AQUABEAM ROBOTIC SYSTEM; FLUID JET REMOVAL SYSTEM Back to Search Results
Catalog Number AB2000
Device Problem Visual Prompts will not Clear (2281)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/06/2024
Event Type  malfunction  
Event Description
A male patient underwent an aquablation procedure for symptomatic benign prostatic hyperplasia (bph).Procept biorobotics corporation became aware that following the first treatment pass, the aquabeam robotic system generated an "e22 - motorpack error" and "e23 - motorpack error" which were cleared with some troubleshooting steps.The aquabeam robotic system then generated an "e31 - motorpack error" which could not be cleared.As a result a second aquabeam handpiece was opened and procedure was completed successfully.The reported event caused a surgical delay of over 20 minutes.There were no adverse health consequences to the patient due to this event.
 
Manufacturer Narrative
Root cause of reported event has not yet been established.Investigation by manufacturer is currently in-process.Submission of this report does not constitute an admission that the manufacturer's product caused or contributed to the event.
 
Manufacturer Narrative
The aquabeam handpiece was not returned for investigation of the reported event.The root cause of the reported event was unable to be established as the device was not returned for investigation.A review of the device history record (dhr) for aquabeam robotic system/serial number (b)(6) and the aquabeam handpiece / lot number 24c11890 was performed, which confirmed that there were two (2) non-conformances issued to this lot during the manufacturing process that could potentially be related to the reported event.The affected units within the lot were segregated and reworked to address the non-conformances.The handpiece passed final inspection prior to release for distribution.The aquabeam robotic system user manual, um0101 rev.F, states the following: table 5: system detected errors and faults.E22 - motorpack error: release foot pedal and click x.1) if error persists, reconnect handpiece to motorpack.2) if error continues, replace handpiece.E23 - motorpack error: release foot pedal and click x.1) if error persists, reconnect handpiece to motorpack.2) if error continues, replace handpiece.E31 - motorpack error: release foot pedal and click x.If error persists, replace handpiece.Submission of this report does not constitute an admission that the manufacturer's product caused or contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AQUABEAM ROBOTIC SYSTEM
Type of Device
FLUID JET REMOVAL SYSTEM
Manufacturer (Section D)
PROCEPT BIOROBOTICS CORPORATION
150 baytech dr
san jose CA 95134
Manufacturer (Section G)
PROCEPT BIOROBOTICS CORPORATION
150 baytech dr
san jose CA 95134
Manufacturer Contact
doria esquivel
150 baytech dr
san jose, CA 95134
6502327291
MDR Report Key18951218
MDR Text Key338364749
Report Number3012977056-2024-00083
Device Sequence Number1
Product Code PZP
UDI-Device IdentifierB614AB20001
UDI-Public+B614AB20001/16D20230822L
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
DEN170024
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/03/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/21/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberAB2000
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/29/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured08/22/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient SexMale
-
-