It was reported that the device was beeping, and the patient was going to be seen as a result for a follow up.Provocation testing and x-ray was suggested to be done as this right ventricular (rv) lead had been implanted for seventeen years.Technical services (ts) reviewed the data and noted that over the past year the average shock impedance measurements were greater then 160 ohms and increasing to 180 ohms.The rv pace thresholds had remained constant and review of the available electrograms (egm) was clean of artifacts with no recent therapy.Ts noted that observations of gradually rising shock impedance can be attributed to patient related factors resulting in calcification/mineralization buildup on the coil.Ts wanted to assure the effectiveness of shock delivery and noted that when the average shock impedance for a single coil lead is above 150 ohms, technical service recommends consideration be given to a commanded 1.1j r-wave sync shock.If the value after the commanded shock is less then 120 ohms regular follow up was recommended, and if the value is greater then 120 ohms to check if 145 ohms is reached further investigation would be warranted.The field representative noted that the superior vena cava coil was turned off and the values had decreased to less than 145 ohms.At this time the rv lead remains implanted.No adverse patient effects were reported.
|