It was reported that the patient had apparently fallen asleep at home and had a pump stop that was presumed to be secondary to loss of battery power on (b)(6) 2024, but could not be confirmed.The patient stated that they did reconnect themself to battery power.The healthcare provider's interrogation dated back to (b)(6) 2024 so they were not able to see the patient's reported event and log files were submitted for review to confirm the sequence of events.Log files did not capture any pump stop, as it may have ben purged by clusters of pulsatility index (pi) events on (b)(6) 2024 to (b)(6) 2024.The log files did captured a few low power advisories due to battery depletion on (b)(6) 2024.Otherwise, there were no notable alarm condition or parameter changes.Based on the data visible to us in the log file the mechanical circulatory support (mcs) equipment was operating as intended electronically and mechanically.
|