Model Number 97715
Device Problems Energy Output Problem (1431); Electromagnetic Compatibility Problem (2927)
Patient Problem Insufficient Information (4580)
Event Date 02/29/2024
Event Type  malfunction  
Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event Description
Information was received from multiple sources (patient, manufacturer representative, healthcare provider) regarding a patient who was implanted with an implantable neurostimulator (ins).It was reported that there was interference with the ventilator.Message left on the rep's voice mail from patient stating, "i've been a spinal stimulator patient at the pain clinic.I had been put on a ventilator for approximately 2 years.The ventilator mask and ventilator system had magnets init, and it created a magnetic field.It would impact my stimulator.I didn't realize what was going on.But, anyway, i got notice from ventilator company that this would happen.And no one warned me about it prior to that.I'm probably going to have to change my ventilator system completely so there's no magnetic field.I wanted to let you know this, since my spinal stimulator was impacted by my ventilator system.The name of it is astral 150 resmed ventilator and mask system.The mask had magnets in it and it created a magnetic field and it did impact my stimulator.Please put this in my record so you will have a record of it.The patient called patient services and inquired about the potential for damage to the devices and the magnetic field that could cause them to have issues.Agent reviewed compatibility information with caller.Caller mentioned they have lung failure and have to use the ventilator due to that.Caller mentioned that on and off stimulation has felt different than before and  that they think that the ventilator has affected their stimulation.Caller reported it doesn't seem like the same pulse and/or sensation for probably 2 years.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTELLIS
Type of Device
STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key18951696
MDR Text Key338708574
Report Number3004209178-2024-07532
Device Sequence Number1
Product Code LGW
UDI-Device Identifier00643169781702
UDI-Public00643169781702
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P840001
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 03/21/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/21/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/28/2019
Device Model Number97715
Device Catalogue Number97715
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/29/2024
Date Device Manufactured05/10/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age73 YR
Patient SexMale
-
-