Philips received a complaint on the efficia dfm100 defibrillator indicating that the power supply exploded causing damage.There was no patient involvement.The fse evaluated the device on site and determined this was a malfunction of the power module.The fse found that the "explosion" caused damage to the i/o pca and the therapy assembly as well.The fse replaced the power module, i/o pca, and therapy assembly resolving the reported issue.Based on the information available and the testing conducted, the cause of the reported problem was a malfunction of the power module.The reported problem was confirmed.The fse replaced the power module, i/o pca, and therapy assembly resolving the reported issue.It has been concluded that no further action is required at this time.The efficia dfm100 defibrillator, model# 866199, is substantially similar to the heartstart xl+ defibrillator (model # 861290) and will be reported in the united states under device model # 861290.
|