• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOSENSE WEBSTER INC THERMOCOOL® SMART TOUCH® SF BI-DIRECTIONAL NAVIGATION CATHETER; CARDIAC ABLATION PERCUTANEOUS CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOSENSE WEBSTER INC THERMOCOOL® SMART TOUCH® SF BI-DIRECTIONAL NAVIGATION CATHETER; CARDIAC ABLATION PERCUTANEOUS CATHETER Back to Search Results
Catalog Number D134801
Device Problem Patient Device Interaction Problem (4001)
Patient Problem Cardiac Tamponade (2226)
Event Date 02/27/2024
Event Type  Injury  
Event Description
It was reported a patient underwent a premature ventricular contraction cardiac ablation procedure with a thermocool® smart touch® sf bi-directional navigation catheter and the patient experienced cardiac tamponade treated with a pericardiocentesis after the procedure was completed.The patient required extended hospitalization about 1 week for follow-up observation.Atrial septal puncture was not performed.Ablation was performed prior to the cardiac tamponade identified.Steam pop was not observed.Blood pressure recovered and patient has fully recovered.Physician's opinion on the cause was patient condition.Correct settings were used on the devices and no error messages appeared on the equipment.
 
Manufacturer Narrative
E1 initial reporter phone: (b)(6).Since no device has been received for analysis, no product investigation can be performed, and the customer complaint cannot be confirmed.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster, inc., or its employees that the report constitutes an admission that the product, biosense webster, inc.Or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Manufacturer's reference number: (b)(4).
 
Manufacturer Narrative
On 28-mar-2024, the product investigation was updated with the manufacturing record evaluation.A manufacturing record evaluation was performed for the finished device number lot 31210591l and no internal action related to the complaint was found during the review.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THERMOCOOL® SMART TOUCH® SF BI-DIRECTIONAL NAVIGATION CATHETER
Type of Device
CARDIAC ABLATION PERCUTANEOUS CATHETER
Manufacturer (Section D)
BIOSENSE WEBSTER INC
31 technology drive, suite 200
irvine CA 92618
Manufacturer (Section G)
BIOSENSE WEBSTER INC (JUAREZ)
circuito interior norte
1820parque industrial salvacar
juarez 32599
MX   32599
Manufacturer Contact
kate karberg
31 technology dr
irvine, CA 92618
3035526892
MDR Report Key18952112
MDR Text Key338302803
Report Number2029046-2024-00969
Device Sequence Number1
Product Code LPB
UDI-Device Identifier10846835010145
UDI-Public10846835010145
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
P030031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/21/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue NumberD134801
Device Lot Number31210591L
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/28/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured12/13/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
SMARTABLATE GENERATOR; UNK CARTO
Patient Outcome(s) Required Intervention; Hospitalization; Life Threatening;
Patient Age72 YR
Patient SexMale
Patient Weight67 KG
-
-