• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENDOLOGIX SANTA ROSA OVATION IX; ILIAC LIMB

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENDOLOGIX SANTA ROSA OVATION IX; ILIAC LIMB Back to Search Results
Model Number TV-IL1414140-J
Device Problem Unintended Movement (3026)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/07/2024
Event Type  Injury  
Event Description
The patient was treated for an abdominal aortic aneurysm (aaa) on (b)(6) 2023, with the implant of an alto abdominal aortic graft system.Endologix was made aware on (b)(6) 2024, that the right limb was pulled back slightly.There was no endoleak.Reintervention was completed on 08 march 2024, with the implant of an additional ovation ix iliac limb to extend further into the common.Final patient status was not reported.
 
Manufacturer Narrative
The devices involved in this event will not be returned for evaluation and remain implanted in the patient.Patient medical records and imaging studies will be requested for further evaluation by a clinical specialist.If additional information pertinent to the incident is obtained, a follow-up report will be submitted.H3 other text : device remains implanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OVATION IX
Type of Device
ILIAC LIMB
Manufacturer (Section D)
ENDOLOGIX SANTA ROSA
3910 brickway blvd
santa rosa CA 95403
Manufacturer (Section G)
ENDOLOGIX SANTA ROSA
3910 brickway blvd
,
santa rosa CA 95403
Manufacturer Contact
gary kirchgater
3910 brickway blvd
,
santa rosa, CA 95403
8009832284
MDR Report Key18952433
MDR Text Key338268998
Report Number3008011247-2024-00026
Device Sequence Number1
Product Code MIH
UDI-Device Identifier00850007370220
UDI-Public(01)00850007370220(17)260317
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P120006
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional,Company Representative
Reporter Occupation Physician
Remedial Action Other
Type of Report Initial
Report Date 03/07/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/21/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberTV-IL1414140-J
Device Lot NumberFS031623-30
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/07/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/24/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
ALTO MAIN BODY (LN FS020323-29); OVATION IX ILIAC LIMB (LN FS042823-10); OVATION PRIME FILL POLYMER (LN FF041123-02)
Patient Outcome(s) Required Intervention;
Patient Age68 YR
Patient SexMale
-
-