Reportable based on device analysis completed on 04mar2024.It was reported that the coil was difficult to coil up, was difficult to pull, and the coil was stretched.The target lesion was located in the bronchial artery.A 3mm x 6cm interlock was selected for use.After the bleeding point was found by an angiography catheter, a non-boston scientific microcatheter was used to perform the embolization.During deployment of the coil, it was noted that it was difficult to coil up due to the relatively thin blood vessel.The coil was withdrawn into the sheath, and it was difficult to pull, resulting in the coil being stretched.The microcatheter was pulled back and the procedure was completed with another of the same device.There were no patient complications, and patient was stable post procedure.However, device analysis revealed that the zap tip was detached.
|
Device evaluated by mfr: the main coil, the introducer sheath and the pusher wire were returned for evaluation.It was visually observed that the main coil was bent and stretched.Also, the zap tip was detached, and the coil of the pusher wire was unraveled.The functional inspection could not be performed due the main coil and the pusher wire are not interlocking.Dimensional inspection of the main coil revealed the components were within specification.No other issues were identified during the product analysis.
|