The subject device underwent visual and functional inspection and the allegation was confirmed.Images were displayed intermittently, and color bars and flickering occurred.Additionally, camera cable was broken and cable watertightness was defective.The camera cable was broken and the watertightness was defective, indicating that moisture may have penetrated inside the camera.Olympus assume that the charged coupled device inside the camera malfunctioned, which prevented normal communication and resulted in the image abnormality indicated.Based on the investigation, no nonconformances were found.A definitive root cause cannot be identified.A review of the device history record found no deviations that could have caused or contributed to the reported issue.To mitigate risk/harm to the patient and damage to the device, the instructions for use provides the following: precautions for disappeared or frozen endoscopic image before use[warning] ·if the endoscopic image disappears unexpectedly or the frozen image cannot be restored during an examination, immediately stop using the camera head and withdraw the endoscope from the patient, ¿withdrawal when any irregularity is observed¿.Continued use of the endoscope under this condition could result in patient injury, bleeding, and/or perforation.Reprocessing: general policy.[warning].·never clean, disinfect, or sterilize this equipment together with sharp-tipped instruments (tweezers, forceps, knives,etc.).These could scratch or tear holes in the camera cable,which could allow water to penetrate and damage electrical circuits inside the instrument.Olympus will continue to monitor the field performance of this device.Should additional relevant information become available, a supplemental report will be submitted.
|
This supplemental report was submitted to provide additional information from the investigation.The device evaluation also found rust on the coupler.A definitive root cause cannot be identified.This issue is addressed in the instructions for use (ifu) preparation and inspection inspection of the camera head when turning the endoscope mount lock counterclockwise with holding it lightly, confirm that the endoscope mount lock is free from looseness or backlash.When operating the endoscope mount, confirm that the endoscope mount is free from looseness or backlash.
|