ABBOTT MEDICAL MITRACLIP G4 CLIP DELIVERY SYSTEM (MDD); MITRAL VALVE REPAIR DEVICES Back to Search Results
Catalog Number CDS0702-XTW
Device Problems Break (1069); Difficult or Delayed Positioning (1157); Improper or Incorrect Procedure or Method (2017); Difficult to Open or Close (2921)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/29/2024
Event Type  malfunction  
Manufacturer Narrative
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event Description
It was reported that a mitraclip procedure was performed to treat a mixed mitral regurgitation (mr) with grade of 4, posterior flail, prolapsed posterior leaflet, dilated left atrium.The mitraclip system was prepared as per instruction for use (ifu) and functioned as intended.During the procedure, the clip was inverted for re-alignment and attempt to get back into the atrium.Echocardiogram showed that one gripper did not raise and remained at 120 degrees.During re-positioning, the clip was caught in chordae and was able to be freed with standard troubleshooting but was noted to be difficult.It was noted that the cds was curved to approximately 110 degrees.After removal, the clip was inspected, showing that the gripper line was not on the gripper but had been pulled into the system.A replacement device was used to complete the procedure.Two clips were implanted, reducing mr to grade 3.There was no clinically significant delay in the procedure and no adverse patient effects.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MITRACLIP G4 CLIP DELIVERY SYSTEM (MDD)
Type of Device
MITRAL VALVE REPAIR DEVICES
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005070406
3885 bohannon drive
menlo park CA 94025
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key18952612
MDR Text Key338305936
Report Number2135147-2024-01298
Device Sequence Number1
Product Code NKM
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P100009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 03/21/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/21/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberCDS0702-XTW
Device Lot Number30912R1082
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/29/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/13/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
STEERABLE GUIDE CATHETER
-
-