It was reported that there were concerns about this cardiac resynchronization therapy defibrillator (crt-d) threshold as it has lowered from implant on the left ventricular (lv) channel.The boston scientific representative noted that they were worried that the lead may have moved.It was noted that the clinic reprogrammed the parameters for this device.The representative wanted to discuss troubleshooting.Ts noted that the thresholds improved.The device will continue to be monitored.The device remains in use.No adverse patient effects were reported.Additional information received indicates that there were concerns this device exhibited loss of capture (loc) on the lv channel during the threshold test as there were high capture thresholds.Technical services (ts) reviewed the reports of the threshold test and noted that there was an intermittently early lv signal that inhibited the lv pace.Ts suspected that the device oversensed a left atrial signal.Ts suggested reprogramming, which resolved the early lv signal from appearing on the electrogram (egm).The device remains in use.No adverse patient effects were reported.
|